YOKOGAWA ZR402G-M
Product Overview: The ZR402G is the converter (transmitter) unit of the Yokogawa ZR22G/ZR402G split-type zirconia oxygen analyzer system. It receives the signal from a zirconia sensor mounted in a high-temperature flue and converts it into a standard oxygen-concentration output. In the model code, the “-M” suffix denotes the cable entry type (M20 × 1.5 mm). The converter is designed for combustion monitoring and control across energy-intensive industries.
Detailed content
Technical Specifications:
- Manufacturer: Yokogawa Electric Corporation
- Product Type: Converter / transmitter (separate type)
- Measurement Object: Oxygen concentration in combustion exhaust gas and mixed gases
- Measurement System: Zirconia sensor
- Measurement Range: 0.01 to 100 vol% O₂
- Output Signal: 4 to 20 mA DC, two outputs, maximum load resistance 550 Ω
- Settable Range: Any range from 0–5 vol% O₂ to 0–100 vol% O₂ in 1 vol% steps, or partial ranges
- Digital Communication: HART, 250 to 550 Ω depending on field devices on the loop (multidrop mode)
- Sample Gas Pressure: –5 to +250 kPa (pressure compensation option required above +5 kPa)
- Sample Gas Temperature: 0 to 700 °C (probe alone); 700 to 1871 °C with 0.15 m probe, JIS 5K 32 FF flange and ZR22P high-temperature adapter
- Probe Length Options: 0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0, 3.6, 4.2, 4.8, 5.4 m
- Contact Outputs: Four points (one fail-safe, normally open)
- Contact Inputs: Two points
- Calibration: Zero/span calibration; automatic, semi-automatic and manual via LCD touch screen
- Display: 320 × 240 dot-matrix LCD touch screen
- Converter Construction: Non-explosion-proof, outdoor installation, equivalent to NEMA 4X / IP66
- Ambient Temperature: Probe (ZR22G) –20 to +150 °C; converter (ZR402G) –20 to +55 °C
- Power Supply: 85 to 264 V AC, 45 to 66 Hz
- Reference Gas System: Instrument air (preferred) 50 kPa (200 kPa with check valve) plus furnace pressure; consumption approximately 1 Nl/min, natural convection
Performance:
- Repeatability: ±0.5 % of maximum set range
- Linearity: ±1 % of maximum set range
- Drift: ±2 % of maximum set range per month
- Response Time: 90 % response within 5 seconds
Functional Features: LCD touch screen for settings, calibration and oxygen-concentration trend display; on-line replaceable long-life zirconia sensor; field-replaceable heater assembly in the probe; remote maintenance via HART; self-diagnosis and maintenance functions; split-type design separating electronics from the high-temperature probe.
Application Scenarios: Combustion monitoring and control in steel, power generation, petrochemical, ceramics, paper, food and textile industries, and in combustion equipment such as incinerators and small to medium boilers; boiler combustion optimisation, flue-gas monitoring, high-temperature furnace oxygen measurement and emission control.
Related Models: ZR22G split-type zirconia sensor (probe); ZR22S explosion-proof detector; ZR402G-P (G1/2 thread cable entry); ZR402G-T (1/2 NPT thread cable entry).
Installation Requirements: The converter is mounted outdoors or in the control cabinet with IP66 protection; the probe is installed on the flue wall. Probe and converter are connected by a dedicated cable with a maximum distance of approximately 300 m, subject to loop-resistance limits.