YOKOGAWA PH201G-A1*A/TB

Product Introduction:
The PH201G-A1A/TB is a smart pH and ORP (Oxidation-Reduction Potential) analyzer/transmitter from Yokogawa’s FLXA202/FLXA21 series. It provides continuous, real-time measurement of pH and ORP values in industrial process streams. The “G” designation indicates a glass electrode sensor type. The “A1A” output configuration provides 4–20 mA DC + RS-485 (Modbus RTU) communication. The “TB” suffix indicates a terminal box enclosure configuration.

Detailed content

Measurement Principle:

  • pH Measurement: A glass electrode and reference electrode form an electrochemical cell. The voltage difference between the two electrodes is proportional to the hydrogen ion activity (pH) in the solution, following the Nernst Equation: E = E₀ − (2.303RT/F) × pH
  • ORP Measurement: A platinum (Pt) or gold (Au) indicating electrode measures the overall redox potential of the solution relative to the reference electrode, expressed in millivolts (mV)

Technical Specifications:

Parameter Specification
Model PH201G-A1*A/TB
Measured Parameters pH, ORP (mV)
pH Range -2.00 to +16.00 pH
ORP Range ±1500 mV
pH Accuracy ±0.005 pH
ORP Accuracy ±1 mV
pH Resolution 0.001 pH
ORP Resolution 0.1 mV
Output (A1*A) 4–20 mA DC (isolated) + RS-485 (Modbus RTU)
Power Supply 12–36 V DC (2-wire or 4-wire)
Temperature Compensation Automatic (Pt1000 RTD, built-in)
Electrode Type (G) Glass pH Electrode + Ag/AgCl Reference
Process Connection 1/2″ NPT, 3/4″ NPT, or Flange (DIN/JIS/ANSI)
Process Temperature 0°C to +120°C (32°F to +248°F)
Process Pressure Up to 0.6 MPa (87 psi)
Protection Class IP66 / IP67 (with TB terminal box)
Wetted Materials SUS316L, PEEK, PTFE
Communication RS-485 Modbus RTU (9600/19200 bps)
Approvals ATEX, IECEx, FM, CSA, SIL2

Structural Features:

  • Wall-mount or pipe-mount transmitter housing with TB terminal box for field wiring
  • Glass pH electrode with double-junction Ag/AgCl reference for contamination resistance
  • Built-in temperature sensor (Pt1000) for automatic temperature compensation
  • IP66/IP67 rated enclosure suitable for outdoor and washdown environments
  • Large LCD display with backlight for local readout

Performance Parameters:

Parameter Value
Response Time (pH) ≤ 30 seconds (95% of step change)
Response Time (ORP) ≤ 15 seconds (95% of step change)
Input Impedance (pH) ≥ 10¹² Ω
Calibration 2-point or 3-point automatic calibration
Slope Check Yes — automatic electrode health diagnosis
Alarm Outputs 2 relay outputs (optional)
Isolation 1500 V AC (input to output to power)

Key Advantages:

  • Ultra-high accuracy (±0.005 pH) — suitable for critical process control
  • Glass electrode with double-junction reference — extended life in dirty or fouling solutions
  • Built-in temperature compensation — eliminates need for separate temperature transmitter
  • RS-485 Modbus communication — direct DCS/PLC integration
  • SIL2 functional safety certified — suitable for safety-instrumented systems
  • Terminal box (TB) enclosure — protects wiring connections in harsh environments

Applicable Industries:

  • Water and Wastewater Treatment
  • Chemical and Petrochemical Processing
  • Pharmaceutical Manufacturing
  • Food and Beverage
  • Pulp and Paper
  • Semiconductor Manufacturing (Ultrapure Water)
  • Mining and Metallurgy

Installation Requirements:

  • Install in a location accessible for electrode maintenance and calibration
  • Ensure adequate flow velocity (> 0.3 m/s) past the sensor to prevent stagnant zones
  • Mount the electrode at least 300 mm below the liquid surface (submersion type) or use appropriate immersion assembly
  • Use shielded cable for RS-485 communication — ground shield at one end
  • Connect power supply with correct polarity — reverse polarity will damage the transmitter
  • Provide adequate drainage if installed outdoors to prevent water ingress into the terminal box

Usage Precautions:

  • Glass electrodes are fragile — handle with extreme care; never touch the glass bulb
  • Store the electrode in pH 4 or pH 7 buffer solution when not in use — never let it dry out
  • Calibrate at least every 2 weeks in critical applications; weekly in dirty process streams
  • Clean the electrode regularly with mild acid (HCl) or pepsin solution for protein fouling
  • Replace the reference electrolyte as per maintenance schedule (typically every 3–6 months)
  • Do not exceed +120°C process temperature — glass electrode will degrade
  • Avoid measuring pH in non-aqueous solvents unless specifically rated for that application
  • Do not subject to mechanical shock or vibration beyond rated specifications