Digital guide

You are here:

YOKOGAWA DX2048T-3-4-2

Product Introduction:
The DX2048T-3-4-2 is a 48-channel universal input paperless recorder from Yokogawa’s DX2000 series. It features a 7-inch color touchscreen display with 3-split screen capability, allowing simultaneous monitoring of multiple process variables. It provides 4 channels of 4–20 mA analog output and 2 channels of alarm contact output. Data is recorded to internal memory and removable USB/CF cards for long-term archival.

Detailed content

Technical Specifications:

  • Model: DX2048T-3-4-2
  • Series: DX2000
  • Number of Input Channels48 universal inputs (thermocouple, RTD, voltage, current, frequency, pulse)
  • Display: 7-inch TFT color touchscreen (T)
  • Screen Split: 3-split display (3) — up to 3 independent display groups
  • Analog Output: 4 channels of 4–20 mA (4)
  • Alarm Output2 channels (2) — relay contact (SPDT, 250 V AC / 30 V DC, 3 A)
  • Recording Media: USB flash drive / CF card
  • Communication: Ethernet (100BASE-TX), Modbus TCP/RTU, OPC UA
  • Power Supply: 100–240 V AC, 50/60 Hz
  • Panel Cutout: 96 × 96 mm (1/8 DIN)
  • Dimensions: 96 × 96 × 150 mm (W × H × D)
  • Weight: Approximately 1.2 kg
  • Protection Class: Front panel IP65 (NEMA 4X)

Functional Features:

  • Universal input accepts TC, RTD, mV, V, mA, frequency, and pulse inputs
  • 3-split screen enables simultaneous viewing of trend, bar graph, and numerical display
  • Event recording with timestamp and annotation
  • Password-protected parameter settings (3 access levels)
  • Scheduled data backup to USB/CF card
  • Web server for remote monitoring via standard browser
  • Math function (addition, subtraction, multiplication, division, square root, totalizer)

Performance Parameters:

  • Scan Cycle: 0.2 seconds (all 48 channels)
  • Display Update Rate: 1 second
  • Recording Interval: 0.1 to 60 seconds (configurable per channel)
  • Memory Capacity: 10 million data points (internal) + removable media
  • Accuracy: ±0.1% of span (DC voltage/current), ±0.2% of span (TC/RTD)
  • Isolation: 500 V AC between channels and to ground

Material Composition:

  • Housing: ABS resin front panel, steel chassis
  • Bezel: Aluminum alloy with black anodized finish
  • Terminal Block: Phoenix-type screw terminals, nickel-plated

Structural Features:

  • 1/8 DIN (96 × 96 mm) panel-mount form factor
  • Front-face touchscreen with bezel-mounted membrane buttons
  • Removable terminal cover for wiring access
  • Bottom-loading USB port and CF card slot

Working Principle:
The DX2048T acquires analog and digital signals from field sensors via its universal input modules. The A/D converter digitizes each input at a 0.2-second scan cycle. Processed data is displayed on the 7-inch touchscreen in real time. The recorder continuously writes data to internal flash memory and can be scheduled to archive to USB/CF cards. The 4–20 mA output channels replicate selected input values for transmission to external controllers or indicators. Alarm outputs activate when configured limits are exceeded.

Advantages & Highlights:

  • 48 channels in a single compact 1/8 DIN instrument — high channel density
  • Touchscreen interface simplifies configuration and operation
  • 3-split display provides comprehensive process overview without scrolling
  • Dual recording (internal + removable media) ensures data security
  • OPC UA and Modbus TCP enable seamless integration with DCS and SCADA

Applicable Industries:

  • Oil Refining and Petrochemical
  • Power Plants (boiler, turbine, generator monitoring)
  • Steel and Metal Processing
  • HVAC and Building Automation
  • Water Treatment Plants
  • Food Processing

Installation Requirements:

  • Cut panel opening to 96 × 96 mm with maximum depth clearance of 150 mm
  • Mount in a location free from direct sunlight, excessive vibration, and corrosive atmosphere
  • Use shielded twisted-pair cables for TC/RTD inputs; ground shields at one end only
  • Provide dedicated 100–240 V AC power supply with ≤ 5% voltage fluctuation
  • Ethernet cable: CAT5e or higher, maximum 100 m run

Usage Precautions:

  • Do not exceed the maximum input voltage of 30 V DC or 250 V AC on any channel
  • Ensure proper cold junction compensation for thermocouple inputs
  • Back up data to removable media before performing firmware updates
  • Clean the touchscreen with a soft, dry cloth — do not use solvents
  • Replace the internal battery (for clock backup) every 2 years