Yokogawa DM1075B00
Product Introduction:
The DM1075B00 is a high-precision 6½-digit benchtop digital multimeter manufactured by Yokogawa Test&Measurement. It is designed for laboratory-grade measurements, production testing, calibration workstations, and quality assurance environments requiring the highest level of accuracy and repeatability.
Detailed content
Technical Specifications:
- Resolution: 6½ digits (1,200,000 counts maximum)
- DC Voltage Accuracy: ±0.0015% of reading (basic accuracy)
- AC Voltage Accuracy: ±0.02% of reading (45 Hz to 1 kHz)
- DC Current Accuracy: ±0.012% of reading
- AC Current Accuracy: ±0.05% of reading (45 Hz to 1 kHz)
- Resistance Accuracy (2-wire): ±0.008% of reading
- Resistance Accuracy (4-wire): ±0.003% of reading
- Temperature Input: Thermocouple (J, K, T, E, R, S, B, N) and RTD (Pt100, Pt1000, Cu50, JPt100)
- Frequency Measurement Range: 0.1 Hz to 300 MHz
- Capacitance Measurement Range: 0.1 nF to 100 mF
Functional Features:
- Auto-ranging and manual ranging capability
- Continuous measurement mode with up to 1,000 readings per second
- Trend capture function for monitoring drift over time
- Statistical functions: mean, standard deviation, maximum, minimum, peak-to-peak
- Comparator function with pass/fail judgment and binning
- Relative mode and offset nulling
- Diode test and continuity buzzer
Interface & Communication:
- GPIB (IEEE-488.2)
- RS-232C
- USB 2.0 (USB-TMC compliant)
- LAN (Ethernet, LXI Class C compliant)
Display:
- Large Vacuum Fluorescent Display (VFD) with backlight
- Primary digits: 6½ digits; Secondary digits: 5 digits
Power Supply:
- AC 100 V to 240 V, 50/60 Hz
- Power consumption: approximately 40 VA
Physical Dimensions & Weight:
- Dimensions: 210 mm (W) × 105 mm (H) × 280 mm (D)
- Weight: approximately 3.5 kg
Material Composition:
- Housing: ABS resin with metal chassis
- Front panel: Aluminum alloy with protective coating
- Connectors: Gold-plated binding posts and BNC connectors
Structural Features:
- Benchtop form factor with tilt-stand for optimal viewing angle
- Rubber feet for vibration isolation and stability
- Rear panel houses all I/O connectors in a shielded configuration
Working Principle:
Utilizes a dual-slope integration A/D converter for DC measurements, providing exceptional noise rejection. AC measurements use true RMS conversion. Resistance measurements employ a constant-current source method with 4-wire Kelvin sensing for lead resistance elimination.
Advantages & Highlights:
- Industry-leading 0.0015% DC voltage accuracy
- 6½-digit resolution for ultra-fine measurement granularity
- LXI compliant for seamless integration into automated test systems
- Low thermal EMF design minimizes measurement errors
Applicable Industries:
- Calibration laboratories (NIST/ISO 17025)
- Semiconductor manufacturing test
- Aerospace and defense electronics testing
- Automotive electronics quality control
- Research and development laboratories
Installation Requirements:
- Place on a stable, level, vibration-free benchtop
- Maintain minimum 100 mm clearance on all sides for ventilation
- Ground the chassis via the rear grounding terminal
Usage Precautions:
- Do not exceed the maximum input ratings: 1000 V DC / 750 V AC for voltage; 3 A for current
- Allow 30-minute warm-up period before high-precision measurements
- Use shielded cables for low-level signal measurements to reduce noise pickup
- Do not operate in environments exceeding 80% relative humidity without dehumidification