Yokogawa ANB10D-42M
Product Introduction:
The ANB10D-42M is an intelligent pH/ORP (Oxidation-Reduction Potential) transmitter from Yokogawa’s AN Series, designed for panel mounting. It features a 4-20mA analog output and 2 relay alarm outputs, with built-in temperature compensation. It supports both pH and ORP measurement modes and is widely used in online water quality analysis across water treatment, chemical, food, and other industries.
Detailed content
Technical Specifications:
| Parameter | Specification |
|---|---|
| Measured Parameters | pH / ORP (switchable) |
| pH Range | -2.00 to +16.00 pH |
| pH Accuracy | ±0.02 pH (within 0 to 14 pH range) |
| ORP Range | -1500mV to +1500mV |
| ORP Accuracy | ±2mV |
| Temperature Compensation | Built-in Pt1000 sensor, ATC (Automatic) or MTC (Manual) |
| Analog Output | 4-20mA DC (load ≤600Ω) |
| Relay Outputs | 2 sets (Alarm 1 + Alarm 2) |
| Relay Capacity | 250V AC / 3A, 30V DC / 3A |
| Communication Interface | RS-485 (Modbus RTU, optional) |
| Power Supply | 100-240V AC (50/60Hz) |
| Power Consumption | Approximately 8W |
| Operating Temperature | 0°C to +50°C |
| Protection Rating | IP65 (front panel) |
Material Composition:
- Front Panel: PC (Polycarbonate) + ABS frame
- Electrode Interface: PVDF corrosion-resistant material
- Housing Seal: EPDM rubber gasket
Structural Features:
- 96mm × 96mm standard panel mount size (DIN 96×96 compliant)
- Front panel with 4-digit LCD display
- 4 push buttons for parameter setting and operation
- Top-mounted electrode terminal block and signal output terminals
Working Principle:
The ANB10D-42M measures the potential difference (mV) generated by the pH glass electrode in response to hydrogen ion activity in the solution. The potential signal is converted to a pH value using the Nernst Equation. For ORP measurement, a platinum electrode measures the solution’s oxidation-reduction potential. The built-in temperature sensor detects solution temperature, and automatic temperature compensation is applied using the Nernst temperature coefficient to ensure measurement accuracy across different temperatures. The measured result is converted via A/D conversion to a 4-20mA output signal, while simultaneously comparing against set upper/lower limits to trigger relay alarms.
Functional Features:
- Dual-mode: pH/ORP in one unit, switchable via parameter setting
- Supports 2-point calibration (up to 5-point calibration optional)
- Relay alarms support upper limit / lower limit / deviation three modes
- Supports manual/automatic temperature compensation switching
- Electrode aging diagnosis function prompts electrode replacement
Advantages & Highlights:
- High-precision pH measurement (±0.02pH) meets stringent process requirements
- Panel mount design suitable for local display at the field
- 2 independent relays allow different alarm values to be set
- Simple operation: 4 buttons complete all settings
- Excellent long-term stability: drift less than 0.05pH/24h
Applicable Industries:
- Municipal water/wastewater treatment plant pH monitoring
- Chemical industry acid-base neutralization process control
- Food & beverage cleaning solution pH monitoring
- Power plant boiler feedwater/condensate pH detection
- Semiconductor industry ultrapure water pH monitoring
Installation Requirements:
- Panel cutout dimensions: 92mm × 92mm (DIN 96×96 standard)
- Recommended installation height: 1.2m to 1.8m from ground level (center of instrument)
- Electrode installation: use flow-through or immersion type, ensure electrode bulb fully contacts measured liquid
- Use shielded cable for signal wiring; ground shield at one end only
- Power wiring: use 1.5mm² or larger conductors
Usage Precautions:
- Calibrate pH electrode daily, minimum once per day
- Electrode must be stored immersed in 3mol/L KCl solution; never store dry
- Avoid contact with extreme pH solutions (pH<1 or pH>12) where measurement error increases significantly
- Set appropriate hysteresis for relay alarms to prevent relay chattering
- Electrode service life: approximately 1-2 years, replace periodically